How to deal with scaling on the electrode of electromagnetic flowmeter

Release Time:2022-04-19

How to deal with scaling on the electrode of electromagnetic flowmeter
1. It is recommended to select sharp or hemispherical electrodes that are not easy to adhere, or select replaceable electrodes, etc.
2. Clean the electrode. Different electrode cleaning methods for electromagnetic flow meter
(1) The surface of platinum electrode is seriously polluted to form an oxide film. First polish the electrode surface with toothpaste, then wash it with pure water, and then immerse it in 3.5mol/l potassium chloride solution for 6 hours before use.
(2) For organic oil and oil pollution, first clean the electrode surface with detergent, then clean it with pure water, and then immerse it in 3.5mol/l potassium chloride solution for 6 hours.
(3) For the electrode polluted by inorganic matter, immerse the electrode in 0.1mol/l dilute hydrochloric acid for 30 minutes, wash it with pure water, and then immerse it in 3.5mol/l potassium chloride solution for 6 hours before use.

previous:Connection types of electromagnetic flowmeter

Next:Faults of electromagnetic flow meter(2)

Related News

Applications of Smart Ultrasonic Water Meters

2024-01-25
Applications of Smart Ultrasonic Water Meters

The Future of Smart Water Meters

2023-12-21
The Future of Smart Water Meters

How to select the Right Electromagnetic Flowmeter for Water Treatment

2023-12-06
How to select the Right Electromagnetic Flowmeter for Water Treatment

Selection Criteria for Electrodes in Electromagnetic Flowmeters

2023-12-02
Selection Criteria for Electrodes in Electromagnetic Flowmeters

Choosing the Right Liner for Electromagnetic Flowmeters

2023-11-29
Choosing the Right Liner for Electromagnetic Flowmeters

LoRaWAN Smart Water Metering system

2023-11-08
LoRaWAN Smart Water Metering system